24 Hours Of ECG Detection 3-6-12-lead Multi-parameter Bed-side Patient Monitor For ICU CCU
Product Description
PM-7000D
Multi-parameter Bed-side Patient Monitor
Overview
Application: Adult/Pediatric/Neonatal/Medicine/Surgery/Operating Room/ICU/CCU
Features: Light & handy
Measurement: ECG/ SpO2/NIBP/HR&PR/Temp/Resp
Network: Wire/wireless network & online update
Triplex alarms: Light/Message/Sound (support real voice)
(High/medium/low levels adjustable)
Electrode Disconnection Alarm (Diagram)
Linux Operating System
Easy operation/stable performance
Electrotome/defibrillation tolerance
Restore within 5 seconds after defibrillation
Power Supply:AC & DC
Data Storage
Trend diagram/table: 1000h
NIBP review: 2000 events
Wave review: 48h (optional)
Alarm review: 200 alarm events
Support drug concentration titration analysis
Physical Specification
Dimension & weight
Equipment: 308mm×141mm×273mm; 4.5 kg
Package: 420mm×290mm×370mm; 6.5 kg
Monitor: 12.1” real color LED screen
Battery:
2200mAh
3h operating after full charge
30min operating after low battery alarm
Electrical Power
AC: 100 ~ 240V, 50Hz/60Hz
DC: Built-in rechargeable battery
Environmental Conditions
Temperature:
Working 5 ~ 40°C
Storage -20 ~ 55°C
Humidity:
Working 20% ~ 85%
Storage 10% ~ 95%
Pressure:
Working 70.0 kPa ~ 106.0kPa
Storage 50.0 kPa ~ 106.0kPa
Technique Specification
ECG
5 Leads: RA, LA, LL, RL, V
Lead mode: Ⅰ,Ⅱ, Ⅲ, AVR, AVL, AVF, V
Gain: ×0.25, ×0.5, ×1.0, ×2.0, (error≤5%)
Sweep speed: 12.5mm/s,25mm/s,
50mm/s (error≤±10%)
Bandwidth: Surgery 1 ~ 20 Hz
Monitor 0.5 ~ 40 Hz
Diagnostic 0.05 ~ 130 Hz
Sensitivity: >200 μV P-P
CMRR: Diagnostic: >90dB
Monitor: ≥100dB
Surgery: ≥100dB
Polarization Voltage: ±300mV
Input Impedance: >5 M ohm
Calibration Voltage: 1mV±5%
ST Segment Range:
-2.0mV ~ +2.0mV (unspecific)
-0.8mV~ +0.8mV (accuracy 0.02mV)
Double Insulation Protection
(4000VAC/50Hz)
Defibrillation tolerance
High frequency electrotome tolerance
AC and Drift tolerance
Support heart Pace-making Analysis
13 Arrhythmia Analysis:
Asystole, V-Fibrillation/Tachycardia,
Single/Paired/Multiform PVC, Bigeminy,
Trigeminy, Polycardia, P-0N-T,
Bradycardia, Leaky-wave, Pacemaker
not captured, Pacemaker failure
SpO2 (Masimo)
Measurement range: 0 ~ 100%
Resolution: 1%
Accuracy: 70% ~ 100% ±3 DIGIT
0% ~ 69% (unspecified)
Pulse rate:
Measure/alarm range 20 ~ 300bpm
Resolution 1bpm
Accuracy 20 ~ 300bpm (±3bpm)
NIBP
Method: Oscillometry
Measure mode: Manual, Auto, STAT
Measure interval in auto mode:
1~5/10/15/30/60/90/120/180/240/480min
Measure period in STAT mode: 5min
Pulse rate range: 40 ~ 240bpm
Measure & alarm range:
Adult mode
SYS 40 ~ 270mmHg
DIA 10 ~ 215mmHg
MEAN 20 ~ 235mmHg
Pediatric mode
SYS 40 ~ 200mmHg